Any time a sample is lost, the analyzer will continue monitoring the stream for at least 45 minutes to make sure it doesn’t have sample. This will take care of temporary losses of sample. If the sample is not restored within this period of time the analyzer will finally shut down (go into standby mode).
This was decided because Hach did not want the APA6000 to continue to run when sample is not present. First of all, if the analyzer is completely out of sample and has internally dried up, we do not want the pump to continue to run (even periodically) and destroy the pump seals. The second reason is that the APA6000 will not give correct reading when it has air within the internal fluid system (tubing or pump).
If the analyzer would periodically try to aspirate sample from a port which has no sample, it may require an hour or more before the air is fully purged and the analyzer can start giving reliable readings. The lack of flow from one sample stream will likely affect the readings of the other sample stream due to the air in the tubing and pump.
You have been directed to this answer based on an ID match.
Click here
to see search results of this phrase.
When sequencing two sample streams, why does the APA6000 analyzer shut down (go into standby mode) if only one sample stream is lost?
Document ID
Document ID TE6720Published Date
Published Date 03/05/2019
Question
When sequencing two sample streams, why does the APA6000 analyzer shut down (go into standby mode) if only one sample stream is lost?
Summary
Dual stream APA6000 analyzer in standby mode when one sample stream is lost.
Answer