What is the purpose of the black paint on the LDO sensor caps?

Document ID

Document ID TE6480

Published Date

Published Date 09/16/2020
Print
Question
What is the purpose of the black paint on the LDO sensor caps?
Summary
Purpose of the black paint on the LDO sensor caps
Answer
This black coating on lab and process LDO caps is to prevent stray light from influencing readings.  The caps should be replaced annually or when about 50% of this black coating is missing.  Using the probe in organics or performing aggressive cleaning will deteriorate this coating.