How do you calibrate the analog outputs on sc1000 controller?

Document ID

Document ID TE6405

Version

Version 3.0

Status

Status Published

Published Date

Published Date 06/26/2018
Question
How do you calibrate the analog outputs on sc1000 controller?
Summary
Analog Output Calibration on sc1000 controller
Answer
The following instructions demonstrate the appropriate analog output calibration procedure: 
  1. Connect a digital multimeter to the appropriate terminals (refer to SC1000 user manual for Output Card Connections): 

     
  2. Using the SC1000 touch screen display, access the menu
  3. Select the TEST/MAINT menu and press enter (check mark)
  4. Select the OUTPUT SETUP menu and press enter
  5. Select the appropriate card/module and press enter
  6. Select FUNCTION TEST and press enter
  • Select the appropriate Output (1, 2, 3, or 4) and press enter 
    Select CURRENT OUT, then choose 20mA and measure with your multimeter, what was the reading?  ___________________mA
  • Then choose 4mA and measure with your multimeter, what was the reading?  ___________________mA
  • Select SET FACTOR
    • Enter an adjustment (this is a multiplication factor, so 1.0 = no adjustment or default factor, enter a number >1.0 when the multimeter reads an analog output value lower than 4/20 mA, enter a number <1.0 when the multimeter reads an analog output value higher than 4/20 mA)
  • Repeat A
  • Repeat B and modify the factor until you measure 4.0 and 20.0 mA (or as close as possible) on your multimeter when selected in the CURRENT OUT menu 
These steps can be repeated for any of the other outputs.
 
Additional Details:   These instructions are to be used after verifying the configuration settings (refer to the article for “Configuring an analog output on an SC1000” for additional information).