How is a Lachat Cadmium Column for Nitrate determination replaced?

Document ID

Document ID TE6373

Version

Version 5.0

Status

Status Published

Published Date

Published Date 03/26/2019
Question
How is a Lachat Cadmium Column for Nitrate determination replaced?
Summary
Cadmium column replacement
 
Answer
 To replace a Cadmium Column for Nitrate determination (Product # 50237A):
 
1. Remove the used cadmium column from the manifold.
2. Remove the flangeless fittings. Properly dispose of the used cadmium column
3. Remove the end plugs from the new column and install the flangeless fittings into the column. Tighten the flangeless fittings securely.
4. Install the cadmium column on the manifold.
 

How is a Lachat Cadmium Column for Nitrate determination installed?