Question
What are the cleaning instructions for the IntelliCAL MTC695 Double Platinum Electrode for AT/KF1000 RedOx Titrations?
Summary
Cleaning instructions for MTC695.
Answer
To clean the IntelliCAL MTC695 Double Platinum Electrode for AT/KF1000 RedOx Titrations (Product # MTC69501) follow these steps:
  1. Put enough Nitric Acid Solution, 1:1, 500 mL (Product # 254049) in a beaker with a stir bar so that the electrode wires are fully submerged with room for the stir bar to spin as well.
  2. Select "Maintenance" from the home screen on the titrator.
  3. Select "Clean PtPt Electrode" from the Maintenance list.
  4. Follow the on screen instructions.