What are the steps to make the probe on the Lachat ASX sampler fit through the lower guide plate?

Document ID

Document ID TE6326

Version

Version 3.0

Status

Status Published

Published Date

Published Date 03/26/2019
Question
What are the steps to make the probe on the Lachat ASX sampler fit through the lower guide plate?
Summary
Lachat sampler dual probe guide plate
Answer
For Lachat ASX autosamplers that include a sample probe guide plate, the plate has two holes in it. A screw underneath the probe assembly can be loosened to turn the guide plate 180 degrees. This will put the other hole into position. The larger hole is needed to accommodate the dual probe used with an auto dilutor. The smaller hole is for the smaller diameter single probe.

 This sampler probe guide plate is visible as item 9 below.


 