How are the toggle switches on the right end panels of the QC8500 SII set to designate the appropriate channel?

Document ID

Document ID TE6138

Version

Version 3.0

Status

Status Published

Published Date

Published Date 03/14/2019
Question
How are the toggle switches on the right end panels of the QC8500 SII set to designate the appropriate channel?
Summary
Dip switch setting for channel address
Answer
Here are the toggle switch settings for each channel. The channel end cover needs to be removed to access these.
  • Channel 1: all off (switched to the left)
  • Channel 2: 1 on (right), 2-4 off (left)
  • Channel 3: 1 off (left), 2 on (right), 3-4 off (left)
  • Channel 4: 1-2 on (right), 3-4 off (left)
  • Channel 5: 1-2 off (left), 3 on (right), 4 off (left) 