Does process concentration affect CL10 electrolyte usage?

Document ID

Document ID TE6113

Version

Version 4.0

Status

Status Published

Published Date

Published Date 03/06/2019
Question
Does process concentration affect CL10 electrolyte usage?
Summary
CL10 electrolyte usage versus process concentration.
Answer
In a process with a higher chlorine concentration, the readings on the CL10 will not remain stable for the specified three to six months that is recommended between electrolyte refills.  Higher chlorine, pH, and temperature will use the electrolyte down faster so it will need to be refilled at more frequent intervals than would occur at lower chlorine concentrations.