What happens if the CA610 Fluoride analyzer is calibrated after a calibration warning appears?

Document ID

Document ID TE6075

Version

Version 5.0

Status

Status Published

Published Date

Published Date 04/04/2019
Question
What happens if the CA610 Fluoride analyzer is calibrated after a calibration warning appears?
Summary
CA610 analyzer calibration and warnings.
Answer
If a calibration warning occurs but it is not cleared, and the analyzer later performs a calibration without errors – the previous warning remains in memory.  It is possible to see a “Marg Slope” warning even though review of the calibration reveals an ideal slope value. 
 