Does a K-Kit need to be used on all channels of a 9240 Sodium Analyzer? Can it be assigned to individual channels?

Document ID

Document ID TE6069

Version

Version 4.0

Status

Status Published

Published Date

Published Date 03/06/2019
Question
Does a K-Kit need to be used on all channels of a 9240 Sodium Analyzer? Can it be assigned to individual channels?
Summary
9240 924x Sodium Analyzer K kit cation kit configuration help
Answer
If the K Kit is used for the 9240 multi stream Sodium Analyzer, it must be used for all the streams the analyzer can support. For example if a three stream analyzer is installed, either all three streams will use the K Kit or none of them will be able to. If this is a requirement then additional analyzers will need to be purchased.