Can the TU5200 Benchtop Laser Turbidimeter measure in units of mg/L?

Document ID

Document ID TE6057

Published Date

Published Date 09/13/2018
Print
Question
Can the TU5200 Benchtop Laser Turbidimeter measure in units of mg/L?
Summary
mg/L units on the TU5200 Benchtop Laser Turbidimeter.
Answer
The TU5200 Benchtop Laser Turbidimeter can measure in units of mg/L. For these units to appear, one must for choose the Degree calibration curve and calibrate using appropriate standards (made from Kaolin or Kieselgur). Once calibration is complete, the instrument will report in units of mg/L only (unless a different calibration curve is selected).

Was this answer helpful?

Submitting...
Thank you for your feedback!
There was an error with your submission. Please try again.