How does the exchange device feature work on the sc1000 controller?

Document ID

Document ID TE6025

Version

Version 3.0

Status

Status Published

Published Date

Published Date 06/26/2018
Question
How does the exchange device feature work on the sc1000 controller?
Summary
Exchange device on sc1000 controller.
Answer
The exchange device feature on the sc1000 controller can be used when replacing a sensor with another sensor that has the same model number and device identifier as the original sensor.  The exchange device feature is used to save configurations for the old sensor onto the new sensor.  This includes 4-20 mA output configuration, modbus telegram information, and sensor configuration.  The original sensor should not be deleted before attempting the exchange device feature or it will not work.  Please find the attached instructions to perform the exchange device function on the sc1000 controller.

 