When the TitraLab® AT1000 is turned off, the syringe empties each time. Is there a way to turn off this feature?

Document ID

Document ID TE5978

Version

Version 2.0

Status

Status Published

Published Date

Published Date 07/10/2018
Question
When the TitraLab® AT1000 is turned off, the syringe empties each time. Is there a way to turn off this feature?
Summary
Automatic purge of syringe when TitraLab® AT1000 is turned off
Answer
Yes, this feature can be deactivated. From the Home screen, select 'Settings'. From the 'Settings' screen, select 'Options'. Press the down arrow key to highlight 'Syringe will empty when switched off' and then press the right or left arrow key to deselect this option. Select 'Done' to save the changes.