Will my older digital sensor be compatible with my new sc200 controller?

Document ID

Document ID TE5975

Published Date

Published Date 05/27/2017
Print
Question
Will my older digital sensor be compatible with my new sc200 controller?
Summary
sc200 controller software compatibility with older digital sensors.
Answer
Please see the attached quick reference guide, which shows a list of digital sensors and the software versions that should be compatible with the sc200 controller.  If a sensor has an older version of software than shown in the quick reference guide, there is a chance it will not scan onto the sc200 controller.  In this case, it is advised to update the firmware of the sensor in order to make it compatible with the sc200 controller.  This firmware update can be done using an sc100 controller, a service cable, and the datacom software, or it can be done at our service bench.  For cases of software incompatibility support, please call technical support for assistance at 800-227-4224.