Why would the BODTrak™II show "Test Mode" after being powered on?

Document ID

Document ID TE5932

Published Date

Published Date 09/06/2019
Print
Question
Why would the BODTrak™II show "Test Mode" after being powered on?
Summary
Test Mode on a BODTrak™II
Answer

If the BODTrak™II displays "Test mode", the following instructions should be followed:
  1. Press 'ON' from the "channel display" screen to access the "range selection" screen.
  2. Press '1' and ''6' at the same time to return to User Mode.
  3. Press 'OFF' to return to the "channel display" screen. "Test Mode" will no longer appear on the display.