Question
What is the replacement for the obsolete Product # 08350=A=0002 Polymetron pH sensor?
Summary
Replacement for Product # 08350=A=0002 Polymetron pH sensor.
Answer
There are two possible replacements for the Polymetron Product # 08350=A=0002:

One is the 8350.4 (Product # 08350=C=0004) and the other is 8350.5 (Product # 08350=C=0005).

The correct replacement depends on the sample characteristics. If there is high temperatures and corrosive media, 8350.4 is more suitable.  If there is fouling sample and organics with more particles, 8350.5 is more suitable.