Why is the QBD1200 analyzer fitting leaking?

Document ID

Document ID TE5863

Published Date

Published Date 05/27/2017
Question
Why is the QBD1200 analyzer fitting leaking?
Summary
QBD1200 analyzer fitting leaks
Answer
The ferrule fittings on the QBD1200 need to be tightened using the plastic ferrule torque tool.  They should not be finger tightened.  If the fitting on the bottom drain line leaks and does not cure by tightening using the plastic ferrule torque tool, use a soft wrench to tighten further, taking caution not to damage the threads on the fitting by overtightening.