How is the cotton "Silver trap" applied to an Orbisphere 311xx series sensor?

Document ID

Document ID TE5827

Version

Version 3.0

Status

Status Published

Published Date

Published Date 03/07/2019
Question
How is the cotton "Silver trap" applied to an Orbisphere 311xx series sensor?
Summary
Method for applying the cotton "Silver Trap" to the Orbisphere 311xx series sensor.
Answer



 
 
Twist a piece of medical sterile cotton wool into a fairly tight string.
 
Soak the cotton in oxygen electrolyte, Catalog Number/ Order Code 2959, for at least one hour prior to installation
 



 
 
Push the string down in the space between the anode and central electrodes.
 
Use a tissue to wipe the center electrodes, and the membrane support piece seating, completely clean of dispersed cotton fibers.
 
Be sure all cotton is below the top edge of the anode!
 



 
 
Screw on the membrane support piece and fill with electrolyte, slowly and gently.
 

 
Form some tiny balls of cotton to a size a bit bigger than the holes in the membrane support piece. Use tweezers and push a cotton ball down in each opening, just under the surface of the membrane support piece.
 
Make a final check to see that no cotton fibers protrude above the sensor service. Add, if necessary, a drop of electrolyte on the center electrode and mount the new membrane.
 
Replace the cotton during each sensor service!
 