What is the difference between the DR 3900 and DR 6000 Spectrophotometers?

Document ID

Document ID TE5743

Published Date

Published Date 06/13/2022
Question
What is the difference between the DR 3900 and DR 6000 Spectrophotometers?
Summary
DR 3900 vs. DR 6000
Answer
The DR 6000 spectrophotometer has a UV lamp while the DR 3900 spectrophotometer does not. As a result, there are more programs on the DR 6000 than on the DR 3900.

The DR 6000 has a Carousel Holder 7 x 1 cm (Product # LZV902.99.00002) and a Carousel Holder 5 x 1 inch (Product # LZV902.99.00012), which can be purchased as optional accessories.

The DR 6000 also has optional software applications which are not available on the DR 3900. These include the Brewery Analysis Package (Product # LZV942), the Drinking Water Analysis Package (Product # LZV925), and the Enzymatic Food Analysis Software (Product # LZV943).

The DR 6000 has tighter specifications than the DR 3900. For example, the wavelength accuracy of the DR 6000 is ± 1 nm. The wavelength accuracy of the DR 3900 is ± 1.5 nm. Data sheets for the DR 3900 Spectrophotometer Data Sheet and Datasheet: DR 6000™ UV-VIS SPECTROPHOTOMETER_DOC052.53.25008, as well as a comparison chart are available.