Is there a procedure for regenerating the Lachat cadmium column?

Document ID

Document ID TE5425

Published Date

Published Date 03/28/2019
Print
Question
Is there a procedure for regenerating the Lachat cadmium column?
Summary
Regeneration procedure for cadmium column
Answer
Hach has no official procedure for regenerating the Cadmium Column for Nitrate determination (Product # 50237A). The ISO procedure might be considered to regenerate cadmium (Cd) columns. Depending on the original quality of the cadmium column, this may nor mot be able to revive the column.
 
  1. Change the sample loop to 50 cm.
  2. Place all the manifold reagent lines in DI water.
  3. Place the Cd column in-line with the switching valve
  4. Run DI water through the Cd column.
  5. Set up a run of five injections of 0.1M HCl, followed by 3 injections of 2% CuSO4 . 5H2O (20g/L). The color of the cadmium should visibly change. If one end of the column is darker, the column can be switched around so the CuSO4 enters from the opposite end.
  6. Stop the pump.
  7. Place the buffer line in buffer solution
  8. Restart the pump.
  9. Run buffer through the Cd column for about 10 minutes.
  10. Install the nitrate manifold properly (using the proper sample loop), and do an efficiency test (NO2 and NO3 at level of the high standard) to verify the column is functioning properly.

Was this answer helpful?

Submitting...
Thank you for your feedback!
There was an error with your submission. Please try again.