The specifications for the Water Purification System Universal Filtration Cartridge (Product # 2967100) are:
  • Hydrogen for strong acid cation resin mixed with weak base anion resin on a 50/50 volume basis
  • Cation resin dyed purple - changes to amber upon exhaustion
  • Removes all ions with the exception of Silica and CO2
  • Capacity 1600 grains as CaCO3
  • Effluent pH 6-7, Conductivity 10-20 µS/cm
  • Applications include:
    • Humidity/environmental chambers
    • Battery water
    • Steamers for food industry
    • Laboratories
    • Cooling loops
    • Steam irons
  • 3 1/2" Diameter x 19 3/4" Length
  • 30 psi maximum rated pressure
  • 100 °F maximum temperature
  • Shelf life of 2 years
  • Design flow rate of 7.2 GPH