Is it normal for the DRB200 to overshoot the set temperature?

Document ID

Document ID TE5350

Version

Version 2.0

Status

Status Published

Published Date

Published Date 05/03/2018
Question
Is it normal for the DRB200 to overshoot the set temperature?
Summary
Temperature overshoot on a DRB200.
Answer
It is normal for the DRB200 to overshoot the temperature set into the instrument.  When this occurs, the temperature will go back down to the setpoint; the temperature should not increase to more than 5° or so over the setpoint.The temperature overshoot is a way to accommodate the large temperature drop that occurs when the block is loaded with cold vials. The cold vials take a lot of energy to accelerate their temperature up to 150C.  The DRB200 beeps when it first reaches the set temperature, prior to overshooting it. The temperature overshoot is not displayed.