What are the custom calibration restraints on the TL23 series benchtop turbidimeters?

Document ID

Document ID TE5289

Version

Version 4.0

Status

Status Published

Published Date

Published Date 09/21/2018
Question
What are the custom calibration restraints on the TL23 series benchtop turbidimeters?
Summary
Custom Calibration TL2300 TL2310 TL2350 TL2360 limitation
Answer
the TL2300 and the TL2310 do not have an option for custom calibrations

The constraint for the custom calibration on the TL2350 and TL2360 are:

1. The first point must be less than 0.1 and greater than 0
2. The second point less than or equal to 40 and greater than the first point
3. From the third point, Point[n] - Point[n-1] must be greater than or equal to 60