Question
How does the TU5 Series' System Check work?
Summary
Full explanation of the System Check feature for the TU5 Series Turbidimeters
Answer
Summary
System Check performs three functions for a TU5 series turbidimeter:
Additional Information and Functionality
Frequency
Each functionality of system check has specific turbidity limits above which this functionality will not work. If turbidity of the sample is above any of these thresholds, no error or warning will show and INACTIVE will be displayed for the functionality on the turbidimeter's SIGNALS menu.
System Check and the Automatic Cleaning Module
While the system check feature is capable of providing a warning about the cleanliness level of the vial and the Automatic Cleaning Module can clean the vial, these items work independently from each other and cannot be tied together. If the Vial Clarity warning is showing frequently, the customer can select a shorter cleaning cycle, or if using the shortest cleaning cycle and still seeing the warning, change the wiper type that is being used.
System Check performs three functions for a TU5 series turbidimeter:
- Vial Presence Detection- Senses whether or not a vial is present inside of the analytical unit, enabling automatic readings when the lid is closed on the TU5200 and providing critical information to prevent flooding the TU5300 and TU5400 process turbs
- Condensation Detection- Senses whether condensation is present on the side of the vial and provides a system warning
- Contamination Detection- Senses whether or not the vial is clean; if contamination is detected will provide a system warning
Additional Information and Functionality
Frequency
- TU5200: Performed when the sample compartment's lid is closed.
- TU5300/TU5400: Performed 2 minutes after the process head is closed and per set interval time on the controller. System check can be set to activate at specific intervals greater than one hour. To initiate a full system check on-demand, navigate to SENSOR SETUP>DIAG/TEST>SIGNALS. Once the SIGNALS menu is chosen, the instrument will perform a full system check.
Each functionality of system check has specific turbidity limits above which this functionality will not work. If turbidity of the sample is above any of these thresholds, no error or warning will show and INACTIVE will be displayed for the functionality on the turbidimeter's SIGNALS menu.
- Vial Presence Detection: functions when turbidity is less than or equal to 200 NTU
- Condensation Detection: functions when turbidity is less than or equal to 10 NTU
- Contamination Detection: functions when turbidity is less than or equal to 1 NTU
System Check and the Automatic Cleaning Module
While the system check feature is capable of providing a warning about the cleanliness level of the vial and the Automatic Cleaning Module can clean the vial, these items work independently from each other and cannot be tied together. If the Vial Clarity warning is showing frequently, the customer can select a shorter cleaning cycle, or if using the shortest cleaning cycle and still seeing the warning, change the wiper type that is being used.