What mV values should an Ion Selective Electrode (ISE) measure in a standard solution?

Document ID

Document ID TE5016

Published Date

Published Date 06/13/2022
Question
What mV values should an Ion Selective Electrode (ISE) measure in a standard solution?
Summary
ISE mV values in standard solutions.
Answer
If an ISE electrode is used to measure two standards that are a decade apart in concentration (meaning the concentration values differ by a factor of 10), the difference in mV values will correspond to the magnitude of the theoretical Nernstian slope value.

For ISE electrodes for monovalent ions at 25 °C , this value will be 59.16mV (some literature may show slightly different figures depending on temperature).

For ISE electrodes for divalent ions at 25 °C, this value will be 29.58.

Divide the measured value by the theoretical one and multiply by 100 % to obtain the slope % value.

The expected mV difference for two standards that are not a decade apart can also be calculated. The equation would be:
(Log(Standard Concentration 1) - Log(Standard Concentration 2)) x Theoretical Nernstian slope value

For example if you wanted to calculate the expected mV difference between a 0.5 ppm standard and a 1.0 ppm standard using an ISE for a monovalent ion, common standards to use with a Fluoride ISE which is monovalent, the calculation would be:

(Log(1)-Log(0.5))x59.16= (0-(-0.30)x59.16= 0.3x59.16= 17.8 mV

This information is located in this document: Application note: Diaphragm / liquid junction (APP-ECH-0023)_DOC042.52.20139

Was this answer helpful?

Submitting...
Thank you for your feedback!
There was an error with your submission. Please try again.