Does Hach sell digital cable by the foot without the digital quick connecters on the ends?

Document ID

Document ID TE4793

Version

Version 2.0

Status

Status Published

Published Date

Published Date 12/08/2016
Question
Does Hach sell digital cable by the foot without the digital quick connecters on the ends?
Summary
Bulk digital cable for extending the cable length on digital sensors.
Answer
There are digital extension cables available in bulk length and they have a Turck connector on one end and are tinned on the other.  See the part numbers below.  Digital extension cables are used only with digital sensors or gateways when connecting to any Hach sc digital controller. Maximum cable length without Termination Box is 100 m (328 ft). Maximum cable length with Termination Box (Catalog Number/ Order Code 5867000) is 1000 m (3280 ft). Instructions for the termination box are attached.

Catalog Number/ Order Code 6122900 - 10 m cable
Catalog Number/ Order Code 6122900-030 - 30 m cable
Catalog Number/ Order Code 6122900-060 - 60 m cable
Catalog Number/ Order Code 6122900-100 - 100 m cable
 
Attachments

Was this answer helpful?

Submitting...
Thank you for your feedback!
There was an error with your submission. Please try again.