Is it possible to use a CL17 to measure Chlorine Dioxide?

Document ID

Document ID TE4791

Version

Version 2.0

Status

Status Published

Published Date

Published Date 03/21/2018
Question
Is it possible to use a CL17 to measure Chlorine Dioxide?
Summary
Using the CL17 total chlorine analyzer to measure Chlorine Dioxide.
Answer
It is possible to use the CL17 to read chlorine Dioxide. However, there are a few things to take into consideration:
The ratio of the molecular weights are similar. That value would be 99.45 (MW of ClO 2 )/70.9 (MW of Cl 2 ) = 1.4. So a 1.0 ppm Cl 2  reading on the display would be equivalent to 1.4 ppm ClO 2  (1.0 x 1.4 = 1.4) This conversion/adjustment would need to be performed by the end user or operator of the CL17 as there is no option in the analyzer to automatically compensate for the difference.

Was this answer helpful?

Submitting...
Thank you for your feedback!
There was an error with your submission. Please try again.