The CL10 chlorine sensor manual indicates using chlorine free water in the chemical calibration. What is chlorine free water?

Document ID

Document ID TE4777

Version

Version 3.0

Status

Status Published

Published Date

Published Date 03/06/2019
Question
The CL10 chlorine sensor manual indicates using chlorine free water in the chemical calibration. What is chlorine free water?
Summary
Chlorine free water definition.
Answer
To get chlorine free water, leave a two or three gallon container of water open to the atmosphere overnight to have the chlorine dissipate. It is best to verify there is no longer chlorine present before performing the zero chemical calibration. This can be done with a handheld unit such as a Pocket Colorimeter II. You can also purchase chlorine free water from a local grocery store.