Question
The TSS Portable shows an asterisk next to the AT1. What does this mean?
Summary
TSS Portable shows an asterisk next to the AT1
Answer
Usually this means a manual calibration curve has been entered.
Here are the steps to remove the curve:
Here are the steps to remove the curve:
- Go to MEAS---> STOP Measurment
- Go to MENU---> Go to CALIBRATION---> CURVE C-Tu---> go down to the last point---> RESET
- Go to MEAS---> START Measurement