What is the maximum cable length for the pH module for the SC200?

Document ID

Document ID TE4576

Version

Version 3.0

Status

Status Published

Published Date

Published Date 03/22/2019
Question
What is the maximum cable length for the pH module for the SC200?
Summary
Maximum cable length for the pH module for the SC200
Answer
For the sc200, the Sensor-to-Controller Distance (maximum) for pHD or LCP sensors: 914 m (3000 ft.) pH Combination electrode w/ preamplifier: 300 m (958 ft.) pH Combination electrode w/o preamplifier: 30 m (100 ft.), depending on environment this distance is shorter. The pHD sensor shall have a built-in preamplifier to enable the signal to be transmitted up to 100 m (328 ft.) with standard cabling and up to 1000 m (3280 ft.) with a termination box.