If the DR900 is specified to have a measurement range of 0-2 Abs, why is it able to read higher than 2.0 Abs in single-wavelength and user-program modes?

Document ID

Document ID TE4525

Published Date

Published Date 03/22/2019
Print
Question
If the DR900 is specified to have a measurement range of 0-2 Abs, why is it able to read higher than 2.0 Abs in single-wavelength and user-program modes?
Summary
Measurement range DR900
Answer
The specification relates to an Abs range that will give linear response across multiple instruments. The majority of the calibrations stored on the DR900 do not go higher than 1.5 Abs, but on a given instrument, higher values are possible. They will not be linear much above 2.0 Abs however.

Was this answer helpful?

Submitting...
Thank you for your feedback!
There was an error with your submission. Please try again.