If the DR900 is specified to have a measurement range of 0-2 Abs, why is it able to read higher than 2.0 Abs in single-wavelength and user-program modes?

Document ID

Document ID TE4525

Version

Version 3.0

Status

Status Published

Published Date

Published Date 03/22/2019
Question
If the DR900 is specified to have a measurement range of 0-2 Abs, why is it able to read higher than 2.0 Abs in single-wavelength and user-program modes?
Summary
Measurement range DR900
Answer
The specification relates to an Abs range that will give linear response across multiple instruments. The majority of the calibrations stored on the DR900 do not go higher than 1.5 Abs, but on a given instrument, higher values are possible. They will not be linear much above 2.0 Abs however.