If the lid is open when the 2100Q and 2100Q ISO turbidimeter is powered on, will this affect the calibration?

Document ID

Document ID TE4435

Version

Version 4.0

Status

Status Published

Published Date

Published Date 03/21/2019
Question
If the lid is open when the 2100Q and 2100Q ISO turbidimeter is powered on, will this affect the calibration?
Summary
Calibration with lid open
Answer
2100Q instrument does not perform any zero/dark measurement during startup so this will not be a concern with this instrument. The zero point is set during calibration. The values are stored with the adjustment during calibration. The instrument will not let you calibrate with the lid open.