What are the Operating Cycle Events for the CA610 Fluoride analyzer?

Document ID

Document ID TE4274

Version

Version 5.0

Status

Status Published

Published Date

Published Date 03/26/2019
Question
What are the Operating Cycle Events for the CA610 Fluoride analyzer?
Summary
Operating Cycle Events for the CA610 analyzer.
Answer
Operating cycle events for CA610 analyzer:

1.  Measurement cycle:  5 pump cycles, then delays 200 seconds and takes the measurement

 
2.  Calibration cycle: 90 pump cycles on Standard 1, delays 200 seconds and takes measurement; 90 pump cycles on Standard 2, delays 200 seconds and takes measurement; 90 pump cycles on process sample and then returns to normal measurements
 
3.  Prime Reagent cycle: 90 pump cycles on sample and reagent
 
4.  Prime Std1 cycle:  90 pump cycles on Standard 1 and reagent
 
5.  Prime Std2 cycle:  90 pump cycles on Standard 2 and reagent
 
6.  Prime All: 90 pump cycles on Standard 1 and reagent; 90 pump cycles on Standard 2 and reagent; 90 pump cycles on sample and reagent.
 

Was this answer helpful?

Submitting...
Thank you for your feedback!
There was an error with your submission. Please try again.