What is the reference setting in the UVAS sc sensor configuration menu?

Document ID

Document ID TE4026

Version

Version 3.0

Status

Status Published

Published Date

Published Date 08/17/2018
Question
What is the reference setting in the UVAS sc sensor configuration menu?
Summary
UVAS sc reference option in sensor configuration
Answer
The UVAS sc sensor has a setting in the configuration menu to turn on or off the REFERENCE. This parameter refers to the turbidity compensation that the UVAS makes. The UVAS compensates all of its measurements for turbidity. One of the methods used for verification of the UVAS is the Deutsches Institute fur Normung ( DIN) Method which does not use a turbidity compensation. To compare the two readings the REFERENCE must be turned off. The instrument comes with REFERENCE ON as the default.
 

Was this answer helpful?

Submitting...
Thank you for your feedback!
There was an error with your submission. Please try again.