What can cause an empty cell compartment reading greater than 0.025 NTU on the TL2300 Turbidimeter?

Document ID

Document ID TE3860

Published Date

Published Date 03/14/2019
Print
Question
What can cause an empty cell compartment reading greater than 0.025 NTU on the TL2300 Turbidimeter?
Summary
TL23 empty cell reading exceeds 0.025 NTU.
Answer
The removable filter assembly on the TL23 may become cloudy or contaminated over time and is the most likely cause of high empty cell compartment readings. Environmental conditions will affect the rate of deterioration. Accumulation of dust, dirt, or haze on the filter lens will increase the empty cell reading and affect the ability of the instrument to measure accurately below the empty cell value. To correct this follow the cleaning procedure outlined on page 32 of the user manual. If this does not correct the condition please contact Hach Technical Support to continue troubleshooting and arrange depot repair service as needed.