When may it be better to use plastic instead of a glass sample cell in a Series 5000 Silica analyzer?

Document ID

Document ID TE3574

Version

Version 3.0

Status

Status Published

Published Date

Published Date 03/07/2019
Question
When may it be better to use plastic instead of a glass sample cell in a Series 5000 Silica analyzer?
Summary
Plastic vs glass sample cell in a Series 5000 Silica analyzer
Answer
The plastic sample cell may be better to use than the glass sample cell in cases where the silica concentration is very low and readings are slightly high using a glass sample cell.  If the silica concentration is very low, there is a chance that silica from the glass cell may leach into the cell and be read during the test.