Is there an estimated detection limit for manganese interference in the DPD methods?

Document ID

Document ID TE3395

Published Date

Published Date 04/16/2019
Print
Question
Is there an estimated detection limit for manganese interference in the DPD methods?
Summary
Estimated detection limit for manganese interference with DPD chemistry
Answer
With the Chlorine (Free & Total) Color Disc Test Kit, Model CN-66 (Product # 223101), 0.09 ppm KMnO4 (or 0.03 ppm Mn) is required to produce a detectable result of 0.1 ppm Cl2.

With the low range DPD methods on the colorimeters and spectrophotometers, 0.018 ppm KMnO4 (or 0.006 ppm Mn) is required to produce a detectable result of 0.02 ppm Cl2.