Why is the BODTrak™II displaying time in minutes rather than days?

Document ID

Document ID TE3254

Published Date

Published Date 09/06/2019
Print
Question
Why is the BODTrak™II displaying time in minutes rather than days?
Summary
BODTrak™II shows minutes not days
Answer
The time should be displayed in days; if it is not, follow these steps to restore the correct time display:
  1. From the "Instrument Setup" screen push "OFF" and the channel display screen, which displays the graph, will appear.
  2. Push "ON" and the "Range Selection" screen will appear.
  3. Push the "1" and "6" keys simultaneously.
  4. Push "OFF," then push and hold the "<" and ">" keys simultaneously to go back to the "Instrument Setup" screen to choose the proper test length.