Does glutamic acid contribute to nitrogenous demand in BOD testing?

Document ID

Document ID TE3207

Version

Version 3.0

Status

Status Published

Published Date

Published Date 10/09/2018
Question
Does glutamic acid contribute to nitrogenous demand in BOD testing?
Summary
Nitrogenous demand of BOD
Answer
Yes, glutamic acid will contribute to nitrogenous demand of BOD. If nitrification inhibitor is added, the result will be lower because the NH 2 is not being oxidized to nitrate or nitrite. Per standard methods, adjust (increase) the amount of seed, even with CBOD, to get within the 198 ± 30.5 mg/L.

Note: Glucose, on the other hand, will contribute only a carbonaceous demand.


Why are CBOD results lower than BOD results when testing a BOD standard solution formulated from GGA?

Was this answer helpful?