What is the shelf life for liquid pH buffers?

Document ID

Document ID TE3070

Published Date

Published Date 06/20/2019
Print
Question
What is the shelf life for liquid pH buffers?
Summary
Liquid pH buffer shelf life.
Answer
Shelf life of any Hach product (in reference to the expiration date printed on the package) is of an unopened bottle stored in the conditions also indicated on the packaging.

Hach pH 4 buffer has a shelf life of four years from manufacture date of an unopened bottle.
Hach pH 7 buffer has a shelf life of two years from manufacture date of an unopened bottle.
Hach pH 10 buffer has a shelf life of one year from manufacture date of an unopened bottle.

Shelf life of a bottle once opened is impossible to determine due to environmental variations specific to the customer location and storage standard operating procedures. (Elevation/pressure for example.) Hach recommends that the date the bottle is opened is written on the bottle and that the pH 4 & 7 buffers be used within six months, the pH 10 buffer within three months of opening.

 

Was this answer helpful?

Submitting...
Thank you for your feedback!
There was an error with your submission. Please try again.