At what temperature are the DRB200 Digital Reactor Blocks calibrated?

Document ID

Document ID TE3066

Published Date

Published Date 08/28/2018
Print
Question
At what temperature are the DRB200 Digital Reactor Blocks calibrated?
Summary
Temperature used to calibrate the DRB200 Digital Reactor Blocks
Answer
The DRB200 Digital Reactor Blocks are calibrated at 100 0C. Usually the temperature measured at this point is very accurate. However, the accuracy may deviate at other points along the temperature range. For example, the actual temperature is usually too high at 1500C. Temperature adjustment at another point may force accuracy errors at other points along the range.