How often should I replace an Ammonia ISE's membrane module?

Document ID

Document ID TE2794

Version

Version 3.0

Status

Status Published

Published Date

Published Date 06/12/2018
Question
How often should I replace an Ammonia ISE's membrane module?
Summary
Membrane replacement schedule.
Answer
A membrane module will last from one week to several months depending on the usage and sample matrix. Shifts in potential, measurement drift, poor response or failed calibration are signs that the membrane should be replaced. Also, replace the membrane module if dark spots or discoloration are seen with visual inspection. Avoid excessive handling of membrane modules or direct contact with membrane surface as this can affect the membrane’s hydrophobic properties, causing shorter membrane life. This is true for any Ammonia probe that uses a gas permeable membrane.