Question
Can the IntelliCAL™ LDO101 Luminescent/Optical Dissolved Oxygen Probes be used in an Acetonitrile solution?
Summary
Chemical compatibility of IntelliCAL™ LDO101 probes with Acetonitrile solutions
Answer
No, unfortunately the LDO sensor cannot be used in an Acetonitrile solution. The probe begins breaking down almost immediately in a 30% Acetonitrile solution. Dissolved oxygen concentrations and percent saturation levels will noticeably drop immediately as the sensor degrades. 