How can I sanitize the IntelliCAL™ LDO101 Luminescent/Optical Dissolved Oxygen Probe?

Document ID

Document ID TE2573

Version

Version 4.0

Status

Status Published

Published Date

Published Date 07/17/2018
Question
How can I sanitize the IntelliCAL™ LDO101 Luminescent/Optical Dissolved Oxygen Probe?
Summary
Sanitizing IntelliCAL™ LDO probes
Answer
The LDO probe can be sanitized by soaking it in a dilute bleach solution. Generally 5-10 mg/L of Chlorine, but do not go over 20 mg/L. A general rule of thumb is 1 drop of bleach per liter of water gives around 1-2 mg/L of Chlorine, so 5 drops of bleach per liter should be sufficient.

 

Was this answer helpful?

Submitting...
Thank you for your feedback!
There was an error with your submission. Please try again.