Are the accessory outputs on the 5500sc analyzer on a switch?

Document ID

Document ID TE2422

Version

Version 4.0

Status

Status Published

Published Date

Published Date 05/15/2018
Question
Are the accessory outputs on the 5500sc analyzer on a switch?
Summary
5500sc analyzer and accessory outputs
Answer
No, the accessory outputs on the 5500sc are not on a switch. These outputs are live at all times that power is present at the analyzer, even when the analyzer is turned off. There is a red LED that will be lit near the accessory output connectors to indicate that power is present. It is critical to physically remove power from the analyzer when making connections to the accessory outputs.
More information on the accessory outputs/power out connections can be found in the article  'What are the accessory outputs on the 5500sc analyzer?'
 