How much sample water is required in each vial to be analyzed by the QBD1200 TOC analyzer?

Document ID

Document ID TE2401

Published Date

Published Date 09/04/2018
Print
Question
How much sample water is required in each vial to be analyzed by the QBD1200 TOC analyzer?
Summary
QBD1200 analyzer and sample volume for analysis.
Answer
The amount of water sample required for analysis by the QBD1200 TOC analyzer depends on the number of replicates selected by the user. For water samples with <4,000 ppb TOC, 8mL of water sample is consumed when analyzing each sample from the vials loaded into the autosampler. In addition, the sample line forward-flush performed at the start of each new vial analyzed uses 5 mL of the water sample in the vial. So, for example, if the autosampler was configured to analyze three samples from each vial, then analyzing each vial would consume: (3 x 8 mL) + 5 mL = 29 mL of water/vial.
A maximum of 5 replicants can be measured with a filled up 40 ml sample vial for the auto sampler.