How do I select my trip point on the SP510 analyzer?

Document ID

Document ID TE2386

Version

Version 3.0

Status

Status Published

Published Date

Published Date 03/06/2019
Question
How do I select my trip point on the SP510 analyzer?
Summary
SP510 and trip point selection
Answer
The SP510 reagents can be ordered in different ranges. The alarm trip point selected should be 40 to 50% higher than the normal effluent hardness.
Alarm Trip Point     Buffer         Indicator
0.3 mg/L                 2768549    2794649
1 mg/L                    2768549    2769049
2 mg/L                    2768549    2769149
5 mg/L                    2768549    2769249
10 mg/L                  2768649    2769249
20 mg/L                  2768749    2769249
50 mg/L                  2768849    2769249
100 mg/L                2768949    2769249

Was this answer helpful?

Submitting...
Thank you for your feedback!
There was an error with your submission. Please try again.