What distance away from elbows, bends or pumps in the pipe is recommended for the insertion mounting for the Solitax sc and TSS sc sensors?

Document ID

Document ID TE2381

Published Date

Published Date 03/06/2019
Print
Question
What distance away from elbows, bends or pumps in the pipe is recommended for the insertion mounting for the Solitax sc and TSS sc sensors?
Summary
Solitax sc, TSS sc and insertion mount installation
Answer
It is recommended to install the sensor at least 1.5m (5 feet) or three times the pipe diameter, whichever is greater, downstream of pumps, valves or pipe elbows.