What causes error 4 on 672X analyzer?

Document ID

Document ID TE2373

Published Date

Published Date 03/06/2019
Print
Question
What causes error 4 on 672X analyzer?
Summary
672X and backup battery
Answer
When a 672X shows ERR 4, it means Memory Loss, default values are in use. This means the back up battery for RAM needs to be replaced. It is located on the back of the display board, but the part number is not given in the manual. The part number is 3A1066 and item description is BATT, NICAD, 3.6V, 110MAH, PC